The Colorimetry Research CR-100 is a high quality colorimeter designed specifically to work with virtually any display technology. The CR-100's machined aluminum and steel housing coupled with a completely solid state design ensure excellent durability while keeping the probe incredibly compact and lightweight (just 12 ounces / 340 grams). What sets the CR-100 apart from other colorimeters are its high quality tristimulus filters. Every CR-100 goes through a strict quality control process including measurement of every tristimulus filter with a double monochromator to ensure the following performance benchmarks are attained:
|
Guaranteed
|
Typical
|
Y (Photopic)
|
≤ 2.0%
|
1.4 - 1.8%
|
Xr (Red)
|
≤ 3.0%
|
2.3 - 2.7%
|
Xb (Red under Blue)
|
≤ 1.2%
|
0.8 - 1.0%
|
Z (Blue)
|
≤ 3.5%
|
2.2 - 2.7%
|
The CR-100 can be used with the CRI Application for manual measurement and logging. It is also compatible with FSI's GaiaColor AutoCal for DM and XMP series monitors, LightIllusion's LightSpace CMS*, LightIllusion's ColourSpace*, and Portrait Display's CalMan allowing for full 3D LUT based calibration of Flanders Scientific monitors.
The CR-100 can be used to measure a wide variety of display technologies and utilizes the Four-Color Matrix Method to calibrate the probe for use with specific display types/spectral distributions. The Four-Color Matrix Method, developed by the National Institute of Standards and Technology (NIST), improves the accuracy of tristimulus colorimeters for measurements of color displays and has been verified to be appropriate for cathode rays tubes (CRTs), liquid crystal displays (LCDs), and organic light emitting diodes (OLED) displays. When ordering the CR-100 (without a spectroradiometer) from Flanders Scientific the probe will be equipped with custom matrix matched selections for all current Flanders Scientific monitors in addition to any matrices provided by CRI.
*LightSpace CMS is sold separately. *CalMAN is sold separately. *ColourSpace is sold separately.
CRI Application screenshots below:
Specifications
Wavelength Range
|
380 - 780 nm
|
Luminance Range
|
0.0002 fL to 1500 fL
0.0002 fL @ Signal to Noise Ratio > 10, 20 seconds exposure
0.001 fL @ Signal to Noise Ratio > 10, 0.5 second exposure
0.001 fL @ Signal to Noise Ratio > 100, 20 seconds exposure
1500 fL @ Signal to Noise Ratio > 8000, 1 second exposure
|
Luminance Accuracy
|
� 2 % @ 0.1 fL, 0.4 second exposure
|
Luminance Repeatability
|
0.2 % @ 0.1 fL, 0.4 second exposure
1.5 % @ 0.01 fL, 0.5 second exposure
1.5 % @ 0.001 fL, 20 seconds exposure
|
Chromaticity Accuracy
|
� 0.0015 x, y @ 0.1 fL, 0.4 second exposure
� 0.0015 x, y @ 0.02 fL, 20 seconds exposure
|
Chromaticity Repeatability
|
0.0005 x, y @ 0.1 fL, 0.4 second exposure
0.0005 x, y @ 0.02 fL, 20 seconds exposure
|
Polarization Error
|
≤ 0.1 %
|
Synchronization Frequencies
|
10 - 500 Hz (Custom Synchronization from 10 Hz to 10 KHz available)
|
Exposure Time Range
|
1 ms to 20 seconds
|
Power Requirements
|
5V, 120 mA (600 mW) via USB 2.0
|
Interface
|
USB 2.0, Ethernet
|
Weight / Mass
|
12oz. (0.34 Kg.)
|
|
|
|
|
|
|
Accessories for this product |
|
|
|