Receptors | Silicon photocells |
---|
Measurement area | Ø27mm (Ø10mm probes also available) |
---|
Acceptance angle | ±2.5° |
---|
Measurement distance | 30±10mm |
---|
Display range | Luminance | 0.0001 to 1000 cd/m2 |
---|
Chromaticity | Displayed as 4- or 3-digit value (selectable) |
---|
Luminance | Measurement range | 0.0050 to 1000 cd/m2 |
---|
Accuracy (white)*1 | 10.00 to 1000 cd/m2: ±2% (0.0050 cd/m2: ±0.0015 cd/m2) |
---|
Repeatability (2σ)*1 | 10.00 to 1000 cd/m2: 0.1% (0.0050 cd/m2: 0.0010 cd/m2) |
---|
Chromaticity | Measurement range | 0.0500 to 1000 cd/m2 |
---|
Accuracy (white)*1 | 120 cd/m2: ±0.002 (± 0.004 for monochrome*2) |
---|
Repeatability (2σ)*1 | 0.0500 to 0.0999 cd/m2: 0.010 2.00 to 1000 cd/m2: 0.001 |
---|
Flicker (Contrast method) | Measurement range | 5 cd/m2or higher |
---|
Display range | 0.0 to 999.9% |
---|
Accuracy | ±1% (Flicker frequency: 30Hz AC/DC 10% sine wave) ±2% (Flicker frequency: 60Hz AC/DC 10% sine wave) |
---|
Repeatability (2σ) | 1% (Flicker frequency: 30Hz AC/DC 10% sine wave) |
---|
Measurement speed (USB) | xyLv | 0.0050 to 0.0999 cd/m2: 4 times/sec. 0.10 to 1.99 cd/m2: 5 times/sec. 2.000 cd/m2to 1000 cd/m2: 20 times/sec. |
---|
Flicker (Contrast) | 16 times/sec. |
---|
Display | Digital | xyLv, XYZ, TΔuvLv, u'v'Lv, RGB analyze |
---|
Analog | ΔxΔyΔLv, R/G B/G ΔG, ΔR B/R G/R |
---|
LCD | 16 characters×2 lines (with backlight) |
---|
SYNC mode | NTSC, PAL, EXT, UNIV, INT |
---|
Measurement subject | Vertical synchronous frequency: 40 to 200Hz |
---|
Memory channels | 100 |
---|
Analyzer function | Standard function |
---|
Interface | USB or RS-232c (38,400 bps maximum) |
---|
Multi-point measurement Max. | 5 points (using optional 4-Probe Expansion Board CA-B15) |
---|
Software Solutions | SDK software (supplied as standard accessory) |
---|
Operating temperature/humidity range | Temperature: 10 to 28°C; Relative humidity 70% or less with no condensation. Luminance change: ±2% ±1 digit of reading for white; Chromaticity change: ±0.002 for white, ±0.006 for monochrome from reading of Konica Minolta's standard LCD*1, 120.0 cd/m2at 23°C 40%RH |
---|
Storage temperature/humidity range | 0 to 28°C: Relative humidity 70% or less with no condensation; 28 to 40°C: Relative humidity 40% or less with no condensation |
---|
Input voltage range | 100 to 240V AC, 50/60Hz, 50VA |
---|
Dimensions (W×H×D) | Main body: 340×127×216mm Probe: Ø49×208mm |
---|
Weight | Main body: 3.58kg; Probe: 530g
|
---|